Suture Strength: Knot-Pull, Straight-Pull and Needle Attachment

USP sets three mechanical requirements for a surgical suture. Knot-pull and straight-pull tensile strength measure the force at which the strand breaks, with a knot tied in it and without one. Needle attachment strength measures the force needed to separate the needle from the strand.

Why it matters

A USP size fixes a diameter range and a minimum strength together, but "minimum strength" is not one number — it depends on how the strand is pulled. A knot concentrates stress at a single point, so a knotted strand breaks at a lower load than the same strand pulled straight. USP therefore normally requires the knotted figure, which is the one that reflects how a suture is actually used. Reading a strength requirement means knowing which of the two it is.

Three requirements, not one

A surgical suture has to survive two different things. The strand has to hold once it is tied, and the needle has to stay on the strand until it is meant to come off. USP tests those separately, and the strand itself is tested two ways.

Knot-pull and straight-pull tensile strength both report the force at which the strand breaks. The difference is whether a knot is tied in the specimen first. Needle attachment strength is a different quantity altogether: not how strong the strand is, but how firmly the needle is joined to it.

The suture sizes chart sets out the other half of the standard — the diameter range at each size. Diameter and strength are fixed together, which is why a size is a classification rather than a measurement.

Knot-pull: the figure normally required

Tie a knot in a strand and it becomes weaker at that point. The strand bends sharply around itself and presses against itself, and both concentrate stress in one place. Pull it and it fails there — at a lower load than the same strand pulled straight.

That is not a defect. It is a property of knotting anything, and it is why USP normally requires the knot-pull figure. A suture is tied in use, so the number that describes it usefully is the one measured on a knotted strand. The knot is tied to a defined procedure — a simple knot, formed the same way every time — so that a result from one laboratory means the same as a result from another.

Straight-pull: where it applies instead

Straight-pull measures an unknotted specimen clamped at both ends. It describes the material and the construction without the knot’s contribution, and USP calls for it in two situations.

The first is where a monograph specifies it. The second is at the fine end of the scale: for sizes finer than 8-0 — metric 0.4 on the sizes chart — tensile strength is measured by straight pull.

There is a third case that follows from the logic rather than from the size. A barbed suture is designed to hold without a knot, and its surface has been cut to do so. Tying a knot on it for a test would place the knot on deliberately altered material, and the resulting figure would describe the test rather than the device. Straight-pull is used instead. Barbed sutures are the clearest case of a device whose specification has to be read on its own terms.

Needle attachment: two different promises

Almost every modern suture is swaged — the strand closed into a bore in the needle’s end, so the two are one unit. USP <871> tests that join by clamping the needle, leaving the swaged portion exposed, and pulling the strand.

What counts as passing depends on what the needle is for.

A standard attachment is not meant to come apart. Its requirement is a minimum, and there is a detail worth noticing: a specimen whose strand breaks before the needle detaches still satisfies it. The join outlasted the thing it was holding, which is exactly what a permanent attachment should do.

A removable attachment is meant to come apart, by a deliberate tug, so the needle can be taken off before the ends are tied. That one cannot be specified as a floor. Too weak and it might separate while the suture is being placed; too strong and it stops doing the job it exists for. So it is held to a range — a minimum and a maximum both — which is the only way to require that something detach reliably and not otherwise.

What the requirements actually give you

The practical value of a floor at every size is that it makes a size mean something across manufacturers. Ordering a 3-0 means ordering a strand within a stated diameter range that has been shown to exceed a stated force. Two products at the same size are not thereby identical, but neither is below the same line.

Three things follow from that, and they are worth keeping distinct:

  • A floor is not a rating. The requirement is a threshold for conformity, not a score. A strand that exceeds its minimum by a wide margin has not become suitable for a different application; it has met the same requirement.
  • Strength is not knot security. Tensile strength is the force at which the strand breaks. Whether a tied knot holds its position is a separate property, belonging to construction, coating and technique.
  • The needle requirement protects a different failure. A strand that meets its tensile floor is no use if the needle comes off in the tissue. Testing the join separately is what covers that, and the removable category is the case where the requirement is about control rather than only about strength.

Finding what a given product meets

This page describes the tests and what they establish. What a particular product was shown to meet is a different question, and it is answered in the instructions for use supplied with that product, against the specification approved in your market. Conformity to gauge, tensile strength and needle attachment is verified batch by batch rather than established once.

Key terminology

Tensile strength
The force at which a strand breaks under tension, reported in kilograms-force or newtons. It is a force, not a pressure, and it is not the same thing as knot security.
Knot-pull tensile strength
Breaking force measured with a knot tied in the specimen. The knot concentrates stress, so the strand fails at the knot and at a lower load than the same strand pulled straight.
Straight-pull tensile strength
Breaking force measured on an unknotted specimen, clamped at both ends. It reflects the material and the construction rather than the effect of a knot.
Needle attachment strength
The force required to separate a swaged needle from its strand. A property of the join rather than of either part on its own.
Standard needle attachment
A needle joined so that it is not intended to be separated. The strand may break before the needle pulls off, and that outcome satisfies the requirement.
Removable needle attachment
A needle intended to be detached deliberately, by a quick tug. It is held to a range rather than a floor, because it must neither come away on its own nor resist coming away on purpose.
USP
United States Pharmacopeia. General chapter <881> covers tensile strength and <871> covers needle attachment; the suture monographs set the limits that apply at each size.

Technical characteristics

Governing chapters
USP <881> Tensile Strength; USP <871> Sutures—Needle Attachment
Strand test normally required
Knot-pull
Where straight-pull applies instead
Where a monograph specifies it, and for sizes finer than 8-0 (metric 0.4)
The knot used
A simple knot, tied to a defined procedure so the result is comparable between laboratories
Needle attachment categories
Standard (not intended to separate) and removable (intended to separate by a quick tug)
Needle attachment limits
A minimum for standard attachment; a minimum and a maximum for removable
The numeric limits
Set out per size in the USP monographs and general chapters — the governing text, and not reproduced here.
What a product actually meets
Stated in the instructions for use supplied with it. Conformity is verified per batch.

How it compares

What each test measures and what a limit on it establishes. No column ranks the tests — they answer different questions about the same device.
TestWhat is measuredHow the specimen is preparedWhat a limit establishes
Knot-pull tensile strengthBreaking force of the strandA knot is tied in the specimen before pullingA floor for the strand as it is actually used — knotted
Straight-pull tensile strengthBreaking force of the strandUnknotted, clamped at both endsA floor for the material and construction, without the knot's effect
Needle attachment (standard)Force to separate needle from strandNeedle clamped, swaged portion exposed, strand pulledThat the join will not give way before the strand does
Needle attachment (removable)Force to separate needle from strandAs above, on a needle meant to detachThat it detaches deliberately, and only deliberately

What clinicians weigh when selecting

Which figure a specification is quoting

Knot-pull and straight-pull are different numbers for the same strand, and the knotted one is lower. A strength figure is only comparable to another figure of the same kind.

A floor is not a rating

USP sets a minimum a strand must exceed at its size. It is a threshold for conformity, not a score, and a strand that exceeds it by more is not thereby suited to a different use.

Strength against knot security

Tensile strength is the force at which the strand breaks. Whether a knot holds its position is a separate property, governed by construction, coating and technique rather than by the tensile requirement.

Size and strength move together

Because the standard raises the strength floor at each larger size, size and minimum strength cannot be chosen independently. Selecting a size is selecting a floor.

Knotless devices

A knot-pull test assumes a strand that is used knotted. Where a device is designed not to be knotted, a knot tied for testing would sit on an altered surface, and the tested value would not describe the device.

What governs in your market

Approved specifications differ by market. The instructions for use supplied with the product in your country is the governing document.

Frequently asked questions

What is the difference between knot-pull and straight-pull tensile strength?

The knot. Knot-pull is measured with a knot tied in the specimen; straight-pull is measured on an unknotted strand clamped at both ends. Both report the force at which the strand breaks, but a knot concentrates stress at one point, so the knotted value is the lower of the two for the same strand.

Which one does USP require?

Knot-pull, normally. It is the figure that reflects how a suture is used, since a suture is tied. Straight-pull is used where a monograph specifies it, and for sizes finer than 8-0.

Why is the knotted value lower?

A knot bends the strand sharply around itself. The bend and the pressure of the strand against itself concentrate stress at that point, so the strand fails at the knot rather than somewhere along its length, and it does so at a lower load than an unknotted specimen of the same strand.

Why are barbed sutures not knot-pull tested?

Because they are not knotted in use, and because their surface has been cut. Tying a knot on a barbed section would test a strand at a point that has been deliberately altered, which does not describe the device. Straight-pull is used instead, against the size the manufacturer designates.

What does needle attachment strength guarantee?

That the join between needle and strand holds. For a standard attachment the requirement is a minimum, and a specimen whose strand breaks before the needle detaches satisfies it — the join outlasted the strand, which is the point.

Why does a removable needle have a maximum as well as a minimum?

Because it is meant to come off. Too weak and it could separate while the suture is being placed; too strong and it defeats the purpose, which is to detach with a deliberate tug so the ends can be tied without a needle on them. A range is the only way to specify both.

Does a higher tensile strength make a suture better?

No. The requirement is a floor a strand must exceed at its size, not a score to maximise. Which material, size and needle are appropriate is a clinical judgement about the case, and the approved indications are stated in the product's instructions for use.

Where are the actual numbers?

In the USP monographs and general chapters, which set the limits per size and per material class. They are the governing text and are not reproduced here. What a given product meets is stated in its instructions for use.

Instructions for use

The instructions for use supplied with the product is the controlled document for the specifications a given product meets. Where this article and the instructions for use differ, the instructions for use govern. The USP monographs and general chapters are the governing text for the test methods and limits described here. Approved specifications vary by market.

Medical & regulatory notice

This article is general reference information about suture test standards for healthcare professionals. It is not medical advice, not a recommendation for any procedure or patient, and not a substitute for the instructions for use supplied with the product or for clinical judgement. Product availability and approved specifications vary by market. Dolphin Sutures manufactures surgical sutures; nothing here should be read as a claim of clinical benefit or of superiority over another material or product.

Written by Dolphin Sutures Medical content, reviewed before publication
Reviewed by Dolphin Sutures In-house regulatory and quality review
Last reviewed Reviewed on a scheduled cycle

Related reading